Notice: The ASA website will be unavailable beginning Thursday, May 17 from 4 p.m. until noon (CDT) on May 18 due to system maintenance.

>

WEBINARS

RSS
>

MEETINGS / EVENTS

RSS

October 13 - 17 2012, 12:00 AM - 12:00 AM

ANESTHESIOLOGY 2012

>

FDA MEDWATCH ALERTS

RSS

May 16, 2012

Hydromorphone Hydrochloride Recall

Summary:

Hydromorphone Hydrochloride Recall

April 18, 2012

Morphine Sulfate Injection USP, 4 mg/mL (C-II), 1 mL fill in 2.5 mL Carpuject by Hospira, Inc: Recall - May Contain More Than Intended Fill Volume

Summary:

Customer report of two Carpujects syringes containing more than the 1 mL labeled fill volume. Opioid pain medications such as morphine have life-threatening consequences if overdosed. Those consequences can include respiratory depression (slowed breathing or suspension of breathing), and low blood pressure.

March 05, 2012

Cardiac Science Powerheart, CardioVive, CardioLife; GE Responder and Responder Pro; and Nihon-Kohden Automated External Defibrillators (AEDs): Class I Recall - Defective Component

Summary: FDA notified healthcare professionals and medical care organizations of the Class 1 recall of the listed AEDs which contain a component that may fail unexpectedly due to a defect. If the component were to fail during a rescue attempt, the AED may not deliver defibrillation therapy, causing serious adverse health consequences, including death. The unit’s self test may not detect the failure or impending failure of the component.

>

ASA FEATURED PRODUCT

Anesthesiology Continuing Education (ACE) Program

SKU: 30702-12CE

... Read more »

Single Copies, Member Price: $300

Respironics, Inc. Trilogy 100 Ventilators: Class I Recall - Device May Stop Delivering Therapy to Patient

Thursday, January 12, 2012

AUDIENCE: Risk Manager, Nursing, Critical Care Medicine


ISSUE: FDA notified health care professionals of the Class 1 recall of this product due to a manufacturing issue can stop delivering therapy to the patient. Part of the blower that circulates air and other gases through the ventilator may move out of position and cause the device to alarm.  Failure to respond could result in the potential for harm or death of a ventilator-dependent patient. See the Recall Notice for a list of affected serial numbers.


BACKGROUND: The Respironics Trilogy 100 ventilator is intended for continuous or intermittent breathing support for the care of pediatric patients, weighing at least 11 pounds, through adult patients who require mechanical ventilation.  The ventilators are used in hospitals, nursing homes and other health care settings and in the home.


RECOMMENDATION: Respironics notified its United States distributors, providers, sales personnel and customers of the recall by telephone on October 11, 2011. The company will continue to work with affected customers to arrange for the return or repair of all recalled ventilators.

Read the Medwatch safety alert, including a link to the Recall Notice, at: